Shimadzu GC-2014 Gas Chromatograph
Main content start
General Information
The Shimadzu GC-2014 gas chromatograph is equipped with FID (with methanizer), ECD (Ni63), and TCD detectors, and is set up for analysis of N2O, CH4, CO2, CO, O2, and N2. The detection/quantitation limits of the instrument are as follows:
- N20: 50ppb to 100ppm by ECD
- CO2, CO: 1ppm to 1% by methanizer and FID, 10ppm to 10% by TCD
- CH4: 0.1ppm to 1% by FID
- O2, N2: 100ppm to 20% by TCD
The Shimadzu 2014 is supported by GCSolutions software running on a Dell computer. The instrument utilizes three valves (two 10-ports, and a four port) in order to separate a single sample injection into three separate slugs which are routed to the TCD, FID, and ECD, respectively.
For further information regarding operation of this instrument, please refer to the Users page.